PRODUCTS・SUPPORT

Laboratory Automation System (LAS)

Laboratory Process Automation Modular System

A new concept modular system automates laboratory processes and eliminates waste to improve quality of testing. Flexible configurations meet your unique operational needs. LPAM shapes laboratory testing for the future.

Features

  • Modular system allows more freedom in configuration and scalability
  • Operate standalone or enhance features with the compatible CLINILOG system
  • Verifies sample tube type and remaining sample volume
  • High-speed decapping and aliquoting at 600 samples/hour
  • Front access supplies replacement and disposal

Functions

Optimal loading for your laboratory operation

Optimal loading for your laboratory operation


Optimal loading for your laboratory operation

Loads 300 or 600 samples at a time with 50-position racks. STAT samples are loaded with 5-position racks for priority processing.

High-speed decapping

High-speed decapping

Features high-speed decapping function capable of processing 600 samples/hour. (rubber caps, over caps, screw caps) Splashproof cover contains sample splashes.

Built-in sample identification function

Built-in sample identification function

L-Vision uses the latest imaging technology to instantly identify tube types, cap colors and remaining sample volume. Captured images are stored to support laboratory quality improvement.

Automated disposal of sample tube caps

Automated disposal of sample tube caps

Disposes decapped sample tube caps into the waste container which can be easily accessed from the front of the system.

Advanced Aliquoting

高機能な分注

Throughput 600 samples/hour (1:1 aliquot). Available for serum and plasma sample aliquoting, detecting the boundary surface of separating agents and blood cells by L-Vision.

Enhanced sorting and unloading area

Enhanced sorting and unloading area

Maximum capacity of 620 samples (50-position rack x 12, 5-position rack x 4). Racks are sorted by destination with up to 64 preset areas per module and can be added as needed.

Flexible configurations automates testing processes

The one-stop-operating system performs loading to sample identification, decapping, sorting, aliquoting, sample transfer to the track system, buffering, and unloading. The modular system with an excellent degree of freedom in configuration and scalability caters to your laboratory environment and optimizes operation.

Flexible configurations automates testing processes Flexible configurations automates testing processes

Standalone operation as a decapper and sorter module.
Or connect the system to a laboratory automation system and operate as a sample loading module.

Flexible configurations automates testing processes Flexible configurations automates testing processes

Dedicated sample unloading module for the laboratory automation system.
Enhance sample loading capacity by adding modules based on the number of samples.

Flexible configurations automates testing processes

Easily connectable to analyzers that require rack loading with rack transfer connection.
Racks are sent through an exclusive exit on the side of the module to the STraS Smart Line.

Specifications

Laboratory Process Automation Modular System [LPAM]
Model: LC
Name LPAM System Controller
Power supply / Power consumption Single phase AC100-240V ± 10%, 50/60Hz, Max. 3,100VA
(100VA for LC alone)
Dimensions / Weight W350 x D1,355 x H1,464mm / 123kg
(excludes LPAM controller and UPS)
Model: F2
Name LPAM Flex In/out 2-drawer
Throughput 1,000 samples/hr
Max. capacity of sample loading 310 samples (50-position rack x 6, 5-position rack x 2)
Power supply / Power consumption Single phase AC100-240V ± 10%, 50/60Hz, Max. 230VA
Dimensions / Weight W470 x D1,355 x H1,464mm / 186kg
Model: F4
Name LPAM Flex In/out 4-drawer
Throughput 1,000 samples/hr
Max. capacity of sample loading 620 samples (50-position rack x 12, 5-position rack x 4)
Power supply / Power consumption Single phase AC100-240V ± 10%, 50/60Hz, Max. 250VA
Dimensions / Weight W720 x D1,355 x H1,464mm / 249kg
Model: DC
Name LPAM Decapper
Throughput 600 samples/hr (rubber caps)
Applicable sample tubes
(types and sizes)
Rubber caps, over caps, screw caps
Outer diameter: 13mm, 16mm; Length: 100mm, 75mm
*Glass tubes are not supported.
Power supply / Power consumption Single phase AC100-240V ± 10%, 50/60Hz, Max. 250VA
Dimensions / Weight W472 x D1,355 x H1,464mm / 171kg
Model: AQ
Name LPAM Aliquoter Module
Throughput 600 samples/hr (500μL aspiration, 1:1 aliquot)
Power supply / Power consumption Single phase AC100-240V ± 10%, 50/60Hz, Max. 600VA
Dimensions / Weight W850 x D1,355 x H1,464mm / 400kg
Supplies (Secondary tubes) 600 pcs / cartridge
Supplies (Aliquot tips) 1,000 pcs / cartridge