Home >> Products >> Electrolyte Analyzer


Electrolyte Analyzer [EA Series]
Electrolyte Analyzer EA07
Electrolyte Analyzer EA08M




Electrolyte Analyzer [EA07]
EA07Secures accurate and stable data by ion selective electrodes that possess high sensitivity and durability.
The very creation of our leading knowledge and technology.
  • Good for serum, plasma, urine and samples from dialysis patients
  • Excellent measurement time - Serum: 24 sec/sample, Urine: 36 sec/sample
  • Crown ether membrane used to measure Na and K
    Molecular Oriented membrane used to measure Cl
Specification
Assays Na, K, Cl
Principle Ion Selective Electrode/ Flow Through, Indirect Method
  Na: Crown Ether membrane, K: Crown Ether membrane
  Cl: MO membrane
Samples Serum, Plasma, Urine, Spinal Fluid, Dialysis Sample
Sample volume 27 µL
Throughput Serum: 150 samples/h (24 sec/sample)
Urine: 100 samples/h (36 sec/sample)
Capacity of sample carousel 28
Control position 6
STAT sample position 2
Measurement range
(mmol/L)
Serum mode - Na: 100-200, K: 1.0-10.0, Cl: 50-200
Urine mode - Na: 10-400, K: 2.0-300, Cl: 15-400
Power supply AC100-220V±10%, 50/60 Hz, 200VA or lower
Dimension 280 (W) x 576 (D) x 547 (H) mm
Weight Approx. 25 kg
Notification Number in Japanese Pharmaceutical Affairs Law: 14B3X000010000BW


Electrolyte Analyzer [EA08M]
EA08MHigh degree of perfection accomplished with the fusion of technological advancement in electrolyte membrane and hardware. EA08M is a high-quality finished module that is an accumulation of our technology cultivated over the years.
EA08M is a LAS connectable type electrolyte analyzer from our long-running EA series.
  • Ion Selective Electrode Method
  • High throughput of 200 samples/hr
  • Small sample volume
  • Connectable to A&T's CLINILOG and other Laboratory Automation Systems
Specification
Assays Na, K, Cl
Principle Ion Selective Electrode/ Flow Through, Indirect Method
  Na: Crown Ether membrane, K: Crown Ether membrane
  Cl: MO membrane
Samples Serum, Plasma, Urine
Sample volume 27 µL
Throughput 200 samples/hr, 150 samples/hr (Urine mode)
Internal sampler Holds 3 Samples
Standard solution Set up in internal sampler
Measurement range
(mmol/L)
        Serum         Urine
Na  100 - 200    10 - 400
K    2.5 - 10        2 - 300
Cl    50 - 200     15 - 400
Power supply AC100-240V, 50/60 Hz, 200VA
Dimension 450 (W) x 630 (D) x 1255 (H) mm
Weight Approx. 55 kg
Notification Number in Japanese Pharmaceutical Affairs Law: 14B3X000010000FC



Home


Products
Laboratory Automation System
Managed Pre-Analysis Module
Electrolyte Analyzer
Blood Coagulation Analyzer
Glucose Analyzer
Laboratory Information System
Data Verification System
Diagnostic Reagents


To the Top